Content deleted Content added
merge two sections |
Ce |
||
(46 intermediate revisions by 25 users not shown) | |||
Line 1:
{{Short description|Microscopy technique}}
'''Imaging particle analysis''' is a technique for making particle measurements using [[digital imaging]], one of the techniques defined by the broader term [[particle size analysis]]. The measurements that can be made include [[particle size]], particle shape (morphology or [[shape analysis (digital geometry)|shape analysis]] and [[grayscale]] or [[RGB color model#Numeric representations|color]], as well as distributions (graphs) of [[Population (statistics)|statistical population]] measurements.▼
== Description and history ==▼
Imaging particle analysis uses the techniques common to [[image analysis]] or [[image processing]] for the analysis of particles. Particles are defined here per [[particle size analysis]] as particulate solids, and thereby not including atomic or sub-atomic particles. Furthermore, this article is limited to [[real image
Given the above, the primary method for imaging particle analysis is using optical microscopy. While [[optical microscope
▲Imaging particle analysis is a technique for making particle measurements using [[digital imaging]], one of the techniques defined by the broader term [[particle size analysis]]. The measurements that can be made include [[particle size]], particle shape (morphology or [[shape analysis (digital geometry)|shape analysis]] and [[grayscale]] or [[RGB color model#Numeric representations|color]], as well as distributions (graphs) of [[Population (statistics)|statistical population]] measurements.
Beginning in the late
▲==Description and history==
▲Imaging particle analysis uses the techniques common to [[image analysis]] or [[image processing]] for the analysis of particles. Particles are defined here per [[particle size analysis]] as particulate solids, and thereby not including atomic or sub-atomic particles. Furthermore, this article is limited to [[real image|real images]] (optically formed), as opposed to "synthetic" (computed) images ([[computed tomography]], [[confocal microscopy]], SIM and other [[super resolution microscopy]] techniques, etc.).
▲Given the above, the primary method for imaging particle analysis is using optical microscopy. While [[optical microscope|optical microscopes]] have been around and used for particle analysis since the 1600's<ref name="Hogg1887">{{cite book|author=Jabez Hogg|title=The Microscope: Its History, Construction, and Application: Being a Familiar Introduction to the Use of the Instrument, and the Study of Microscopical Science|url=http://books.google.com/books?id=wzM5AAAAMAAJ&pg=PA8|year=1887|publisher=G. Routledge and Sons|pages=8–}}</ref>, the "analysis" in the past has been accomplished by humans using the human [[visual system]]. As such, much of this analysis is subjective, or qualitative in nature. Even when some sort of qualitative tools are available, such as a measuring [[reticle]] in the microscope, it has still required a human to determine and record those measurements.
▲Beginning in the late 1800's<ref name="Tissandier1877">{{cite book|author=Gaston Tissandier|title=A History and Handbook of Photography|url=http://books.google.com/books?id=lWo9AAAAYAAJ&pg=PA1|year=1877|publisher=Sampson, Low, Marston, Low, & Searle|pages=1–}}</ref> with the availability of [[photographic plate|photographic plates]], it became possible to capture microscope images permanently on film or paper, making measurements easier to acquire by simply using a scaled ruler on the hard copy image. While this significantly speeded up the acquisition of particle measurements, it was still a tedious, labor intensive process, which not only made it difficult to measure statistically significant particle populations, but also still introduced some degree of human error to the process.
Finally, beginning roughly in the late 1970's, [[Charge-coupled device|CCD digital sensors]] for capturing images and computers which could process those images, began to revolutionize the process by using [[digital imaging]]. Although the actual algorithms for performing [[digital image processing]] had been around for some time, it was not until the significant computing power needed to perform these analyses became available at reasonable prices that digital imaging techniques could be brought to bear in the mainstream. The first dynamic imaging particle analysis system was patented in 1982 <ref>{{US patent|4338024}}</ref>.▼
▲Finally, beginning roughly in the late
As faster computing resources became available at lowered costs, the task of making measurements from microscope images of particles could now be performed automatically by machine without human intervention, making it possible to measure significantly larger numbers of particles in much less time.
== Image acquisition methods ==
The basic process by which imaging particle analysis is carried out is as follows:
# A digital camera captures an image of the [[field of view]] in the optical system.
# A gray scale [[Thresholding (image processing)|thresholding]] process is used to perform [[Image segmentation#Thresholding|image segmentation]], segregating out the particles from the background, creating a [[binary image]] of each particle.<ref name=Gonzalez>{{cite book|last=Gonzalez|first=Rafael C.|title=Digital Image Processing|year=2002|publisher=Pearson Education|isbn=978-8178086293|pages=595–611|author2=Woods, Richard E.}}</ref><ref name="Sankur2004">{{cite journal|last1=Sankur|first1=Bulent|title=Survey over image thresholding techniques and quantitative performance evaluation|journal=Journal of Electronic Imaging|volume=13|issue=1|year=2004|pages=146|issn=1017-9909|doi=10.1117/1.1631315|bibcode = 2004JEI....13..146S }}</ref><ref>{{cite journal|title=A Threshold Selection Method from Gray-Level Histograms|journal=IEEE Transactions on Systems, Man, and Cybernetics|volume=9|issue=1|year=1979|pages=62–66|issn=0018-9472|doi=10.1109/TSMC.1979.4310076|last1=Otsu|first1=Nobuyuki}}</ref>
# [[Digital image processing]] techniques are used to perform [[image analysis]] operations, resulting in morphological and grey-scale measurements to be stored for each particle.<ref name="CarterYan2005">{{cite journal|last1=Carter|first1=R M|last2=Yan|first2=Y|title=Measurement of particle shape using digital imaging techniques|journal=Journal of Physics: Conference Series|volume=15|issue=1|year=2005|pages=177–182|issn=1742-6588|doi=10.1088/1742-6596/15/1/030|bibcode = 2005JPhCS..15..177C |doi-access=free}}</ref>
# The measurements saved for each particle are then used to generate image population statistics,<ref>{{cite web|last=Pouli|first=T.|title=Image Statistics and their Applications in Computer Graphics (2010)|url=http://www.cs.bris.ac.uk/~reinhard/papers/eg2010_tania.pdf|archive-url=https://wayback.archive-it.org/all/20110401055809/http://www.cs.bris.ac.uk/~reinhard/papers/eg2010_tania.pdf|url-status=dead|archive-date=1 April 2011|publisher=Eurographics, State of the Art|accessdate=2 January 2014|author2=Cunningham, D|author3=Reinhard, E.}}</ref> or as inputs to algorithms for filtering and sorting the particles into groups of similar types. In some systems, sophisticated [[pattern recognition]] techniques<ref name="Rosenfeld1981">{{cite journal|last1=Rosenfeld|first1=A.|title=Image pattern recognition|journal=Proceedings of the IEEE|volume=69|issue=5|year=1981|pages=596–605|issn=0018-9219|doi=10.1109/PROC.1981.12027|s2cid=13410801}}</ref><ref>{{cite book|last=Young|first=T. Y.|title=Handbook of Pattern Recognition and Image Processing|year=1986|publisher=Academic Press|isbn=978-0127745602|url-access=registration|url=https://archive.org/details/handbookofpatter0000unse}}</ref> may also be employed in order to separate different particle types contained in a heterogeneous sample.
Imaging particle analyzers can be subdivided into two distinct types, static and dynamic, based upon the image acquisition methods. While the basic principles are the same, the methods of image acquisition are different in nature, and each has advantages and disadvantages.
=== Static imaging particle analysis
Static image acquisition is the most common form. Almost all microscopes can be easily adapted to accept a digital camera via a [[C mount]] adaptor. This type of set-up is often referred to as a [[digital microscope]], although many systems using that name are used only for displaying an image on a [[Video monitor#video display|monitor]].
The sample is prepared on a microscope slide which is placed on the [[Optical microscope#Stage|microscope stage]]. Once the sample has been focused on, then an image can be acquired and displayed on the monitor. If it is a [[digital camera]] or a [[frame grabber]] is present, the image can now be saved in digital format, and image processing algorithms can be used to isolate particles in the field of view and measure them.<ref>{{cite book|last=Russ|first=J.C.|title=Computer-Assisted Microscopy: The Measurement and Analysis of Images|year=1990|publisher=Springer US|isbn=978-1-4612-7868-9}}</ref><ref>{{cite book|author1=Hazelwood, Kristin L.|author2=Olenych, Scott G.|author3=Griffin, John D.|author4=Cathcart, Judith A.|author5=Davidson, Michael W.|editor1=Shorte, Spencer L.|editor2=Frischknecht, Friedrich|title=Imaging Cellular and Molecular Biological Functions|url=https://archive.org/details/imagingcellularm00shor|url-access=limited|year=2007|publisher=Springer|isbn=978-3-540-71330-2|pages=[https://archive.org/details/imagingcellularm00shor/page/n21 3]–43|chapter=Entering the Portal: Understanding the Digital Image Recorded Through a Microscope}}</ref>
In static image acquisition only one field of view image is captured at a time. If the user wishes to image other portions of the same sample on the slide, they can use the X-Y positioning hardware (typically composed of two [[linear stage
The major drawback to static image acquisition is that it is time consuming, both in sample preparation (getting the sample onto the slide with proper dilution if necessary), and in multiple movements of the stage in order to be able to acquire a statistically significant number of particles to count/measure. Computer-controlled X-Y positioning stages are sometimes used in these systems to speed the process up and to reduce the amount of operator intervention, but it is still a time consuming process, and the motorized stages can be expensive due to the level of precision required when working at high magnification.<ref name=Brown>{{cite web|last=Brown|first=L.|title=Dynamic Versus Static Image Acquisition in Particle Imaging|url=http://www.particleimaging.com/dynamic-versus-static-image-acquisition-in-particle-imaging/|work=www.particleimaging.com|accessdate=2 January 2014|archive-url=https://web.archive.org/web/20140103060115/http://www.particleimaging.com/dynamic-versus-static-image-acquisition-in-particle-imaging/|archive-date=3 January 2014|url-status=dead}}</ref>
The major advantages to static particle imaging systems are the use of standard microscope systems and simplicity of [[depth of field]] considerations. Since these systems can be made from any standard optical microscope, they may be a lower cost approach for people who already have microscopes. More important, though, is that microscope-based systems have less depth of field issues generally versus dynamic imaging systems. This is because the sample is placed on a microscope slide, and then usually covered with a [[cover slip]], thus limiting the plane containing the particles relative to the [[optical axis]]. This means that more particles will be in acceptable focus at high magnifications.<ref name=Brown />
=== Dynamic imaging particle analysis ===
[[File:Basic flow through diag on white.png|thumb|Diagram showing flow-through architecture for dynamic imaging particle analysis.]]In Dynamic image acquisition, large amounts of sample are imaged by moving the sample past the microscope optics and using [[flash (photography)#High speed flash|high speed flash]] illumination to effectively "freeze" the motion of the sample. The flash is [[synchronization|synchronized]] with a high [[shutter speed]] in the camera to further prevent motion blur. In a dry particle system, the particles are dispensed from a shaker table and fall by gravity past the optical system. In fluid imaging particle analysis systems, the liquid is passed across the optical axis by use of a narrow flow cell as shown at right.
[[File:
The major drawback to dynamic image acquisition is that the flow cell depth must be limited as described above. This means that, in general, particles larger in size than the flow cell depth can not be allowed in the sample being processed, because they will probably clog the system. So the sample will typically have to be filtered to remove particles larger than the flow cell depth prior to being evaluated. If it is desired to look at a very wide range of particle size, this may mean that the sample would have to be fractionated into smaller size range components, and run with different magnification/flow cell combinations.<ref name=Brown />▼
The major advantage to dynamic image acquisition is that it enables acquiring and measuring particles at significantly higher
▲The major drawback to dynamic image acquisition is that the flow cell depth must be limited as described above. This means that, in general, particles larger in size than the flow cell depth can not be allowed in the sample being processed, because they will probably clog the system. So the sample will typically have to be filtered to remove particles larger than the flow cell depth prior to being evaluated. If it is desired to look at a very wide range of particle size, this may mean that the sample would have to be fractionated into smaller size range components, and run with different magnification/flow cell combinations.
Dynamic imaging particle analysis is used in aquatic microorganism research to analyze phytoplankton, zooplankton, and other aquatic microorganisms ranging from 2 um to 5 mm in size. Dynamic imaging particle analysis is also biopharmaceutical research to characterize and analyze particles ranging from 300 nm to 5mm in size.
▲The major advantage to dynamic image acquisition is that it enables acquiring and measuring particles at significantly higher rates of speed, typically on the order of 10,000 particles/minute or greater. This means that statistically significant populations can be analyzed in far shorter time periods than previously possible with manual microscopy or even static imaging particle analysis. In this sense, dynamic imaging particle analysis systems combine the speed typical of [[particle counter|particle counters]] with the discriminatory capabilities of microscopy.
===Micro-flow imaging===
== References ==▼
''Micro-flow imaging (MFI)'' is a particle analysis technique that uses flow [[microscopy]] to quantify particles contained in a solution based on size. This technique is used in the [[biopharmaceutical]] industry to characterize subvisible particles from approximately 1 μm to >50 μm.<ref>{{cite journal | pmc = 2895433 | pmid=20517661 | doi=10.1208/s12248-010-9205-1 | volume=12 | issue=3 | title=Micro-flow imaging: flow microscopy applied to sub-visible particulate analysis in protein formulations | year=2010 | journal=AAPS J | pages=455–64 | last1 = Sharma | first1 = DK | last2 = King | first2 = D | last3 = Oma | first3 = P | last4 = Merchant | first4 = C}}</ref>
{{Reflist}}▼
▲== References ==
▲{{Reflist}}
[[:Category:Laboratory equipment]]▼
[[:Category:Counting instruments]]▼
[[Category:Microscopy]]
|