The IEC 60601-1 standard requires that two independent precautionary measures (Means of Patient Protection, or MOPPs) are set in place, to protect the patient from electrical shock in the medical system. A network isolator can be constructed to provide one or both required MOPPs. In the case where only a single MOPP is present in a system, a second measure of protection for the medical device must be made. This can be achieved, for example, by installing a fixed, low-resistance earthing cable, bar, or strap, referred to as a potential equalisation conductor. Conformation to the requirements of applicable standards and guidelines, in particular the IEC 60601-1 standard, involves manufacturers themselves declaring [[CE mark|CE]] compliance and/or gaining certification from an approved independent testing laboratory.
|